Labotemp Formaldehyde Autoclave FHA-02 features a 150L chamber capacity, low-temperature sterilization at 60°C and 76°C, Siemens PLC control, and an advanced vacuum-assisted steam formaldehyde sterilization process for heat-sensitive medical devices. Designed for hospitals, research laboratories, endoscopy units, CSSD departments, and healthcare facilities, this formaldehyde sterilizer provides effective sterilization of instruments that cannot tolerate conventional high-temperature steam cycles. The controlled combination of steam, vacuum, and formaldehyde ensures deep penetration into complex instruments, lumened devices, and delicate medical equipment while maintaining material integrity and sterilization reliability.
Engineered for safety, efficiency, and long-term performance, the FHA-02 incorporates a high-precision touch-screen interface, automatic dosing system, intelligent temperature and pressure monitoring, and programmable sterilization cycles. The advanced cleaning and drying function helps minimize formaldehyde residue after sterilization, while the negative-pressure operating system enhances operator safety by reducing leakage risks. Its electric lift door, automatic data recording, network connectivity, and storage of more than 15,000 sterilization records support modern traceability requirements.
Why Choose Labotemp Formaldehyde Autoclave FHA-02?
✓ Low-Temperature Sterilization: Operates at 60°C and 76°C for heat-sensitive instruments
✓ Advanced Vacuum Technology: Enhances sterilant penetration into complex and lumened devices
✓ Complete Traceability: Stores 15,000+ cycle records with printing and network connectivity
| General Specifications | |
| Capacity | 150 L |
|---|---|
| Operating Temperature | 60°C and 76°C |
| Ambient Temperature | 5°C to 40°C |
| Sterilization Time | 2–3.5 hours |
| Sterilant Composition | 2% Formaldehyde, 3% Alcohol, 95% Distilled Water |
| Sterilant Package Size | 2 L |
| Single Cycle Dosage | 0.6–1.3 kg |
| Relative Humidity | ≤85% |
| Atmospheric Pressure | 70–106 kPa |
| Maximum Pressure Resistance | 5 Pa |
| Operating Height | ≤160 cm |
| Sterilization Modes | 60°C Mode, 76°C Mode |
| Drying / Cleaning Function | Automatic |
| Electrical & Construction | |
| Power Supply | AC 380 V, 50 Hz ±1 Hz |
| Power Rating | 5.3 kVA |
| Heating Method | Electric Heating |
| Chamber Material | 5052 Corrosion-Resistant Aluminum Alloy |
| Chamber Wall Thickness | 8 mm |
| Door Material | 5052 Aluminum Alloy, 10 mm Thick |
| Pipe Material | SUS304 Stainless Steel |
| Thermal Insulation | High-Quality Insulation Cotton |
| Loading System | Movable Stainless-Steel Basket |
| Door Type | Automatic Electric Lift Door |
| Steam Generator | Fast Evaporator Type |
| Vacuum Pump | German Water-Ring Vacuum Pump |
| Control & Monitoring | |
| Control System | Siemens PLC Control System |
| User Interface | Industrial Touch Screen |
| Communication Rate | 19.2 Kbps |
| Display Parameters | Temperature, Pressure, Time, Operation Mode, Alarms |
| Temperature Control | PID Control System |
| Temperature Sensor | Embedded High-Precision Sensor |
| Pressure Sensor | High-Precision Built-In Sensor |
| Air Filter Accuracy | 0.22 μm |
| Alarm System | High Temperature Alarm |
| Leakage Protection | Negative Pressure Working System |
| Printing Function | Built-in Printer with Traceability Records |
| Data Storage Capacity | ≥15,000 Sterilization Cycles |
| Consumption & Packaging Information | |
| Energy Consumption | 12 kWh per cycle |
| Water Consumption | 3.95 kg/min |
| Package Size | 1260 × 880 × 1950 mm |
| Package Size (consumables) | 900 × 270 × 530 mm |
| Gross Weight | 420 kg |
FHA-02 is suitable for sterilizing heat- and moisture-sensitive medical instruments, endoscopic devices, lumened equipment, surgical tools, and reusable healthcare accessories in hospitals, CSSD units, endoscopy centers, operating rooms, and research laboratories. Supports traceable sterilization workflows with automated documentation, making it ideal for regulated healthcare environments requiring validated low-temperature sterilization and compliance-focused infection control processes.