Labotemp UV-Vis-NIR Spectrophotometer NSP-05 is a double-beam analytical instrument designed for spectral measurement across ultraviolet, visible and near-infrared regions. With a wavelength range of 190–3200 nm in transmission and specular reflectance modes and 250–2500 nm in integrating sphere mode, this system supports analysis of optical properties in glass, films, materials and other suitable samples. Its Czerny-Turner optical system, deuterium and tungsten light sources, and PMT and PbS detectors support multiple spectral measurement applications.
Featuring automatic calibration, computer-based control and integrated spectral analysis software, NSP-05 provides transmission, absorbance, reflectivity and energy measurement modes. Adjustable spectral bandwidth and sampling intervals support different testing requirements, while dedicated accessories accommodate integrating sphere, powder sample, transmission, thin-film, liquid transmission and reflection measurements. USB 2.0 connectivity enables computer communication, data acquisition and processing for laboratory research and material testing.
Why Choose Labotemp UV-Vis-NIR Spectrophotometer NSP-05?
✓Broad Wavelength Coverage: Spectral measurement from 190 to 3200 nm in specified measurement modes.
✓Double-Beam Optical System: Supports spectral measurement across UV, visible and near-infrared regions.
✓Flexible Sample Testing: Compatible optional accessories for different sample measurement configurations.
| No. | Specification | Value |
|---|---|---|
| General Information | ||
| 1 | Product Type | UV-Vis-NIR Spectrophotometer |
| 2 | Model | NSP-05 |
| 3 | Optical System | Double Beam |
| 4 | Optical Design | Czerny-Turner |
| 5 | Light Sources | Deuterium Lamp, Tungsten Lamp |
| 6 | Detectors | PMT, PbS |
| 7 | Operating Modes | Transmission, Absorbance, Reflectivity, Energy |
| Wavelength Performance | ||
| 8 | Wavelength Range: Transmission and Specular Reflectance | 190–3200 nm |
| 9 | Wavelength Range: Integrating Sphere | 250–2500 nm |
| 10 | Wavelength Accuracy: UV-Vis | ±0.5 nm |
| 11 | Wavelength Accuracy: NIR | ±0.4 nm |
| 12 | Wavelength Repeatability: UV-Vis | ≤0.2 nm |
| 13 | Wavelength Repeatability: NIR | ≤1 nm |
| 14 | Spectral Bandwidth: UV-Vis | 0.1–7 nm |
| 15 | Spectral Bandwidth: NIR | 0.8–20 nm |
| 16 | Available Sampling Intervals | 0.1, 0.2, 0.5, 1, 1.5, 2, 5 and 10 nm |
| Photometric Performance | ||
| 17 | Photometric Range | 0–3 A |
| 18 | Photometric Accuracy | ±0.3%T (0–100%T) |
| 19 | Photometric Repeatability | ≤0.2%T |
| 20 | Stray Light | ≤0.2%T at 360 nm and 420 nm |
| 21 | Stability | ±0.002 A/h at 500 nm, 0 A |
| 22 | Baseline Flatness: 190–3200 nm | ±1%T after 30 minutes warm-up |
| 23 | Baseline Flatness: 200–2500 nm | ±0.5%T after 30 minutes warm-up |
| Optical Components | ||
| 24 | Diffraction Grating: UV-Vis | 1200 lines/mm |
| 25 | Diffraction Grating: NIR | 300 lines/mm |
| Software and Connectivity | ||
| 26 | Computer Interface | USB 2.0 |
| 27 | Software | Operational Software |
| 28 | Automatic Calibration | Supported |
| 29 | Data Acquisition | Automatic |
| 30 | Data Processing | Supported |
| 31 | Data Import and Export | Supported |
| 32 | Compatible Operating Systems | Windows XP, Windows 7 |
| Electrical Requirements | ||
| 33 | Power Supply | AC 110/220 V |
| 34 | Frequency | 50/60 Hz |
| Dimensions and Weight | ||
| 35 | External Dimensions (W × D × H) | 760 × 605 × 300 mm |
| 36 | Package Dimensions (W × D × H) | 870 × 750 × 505 mm |
| 37 | Net Weight | 48 kg |
| 38 | Gross Weight | 60 kg |
Glass Testing: Spectral analysis of glass transmission and reflectance properties.
Energy Efficiency Testing: Measurement of optical properties relevant to energy-efficient materials and glazing.
Construction Material Inspection: Optical testing for construction engineering quality assessment.
Automotive Glass Testing: Spectral measurement of suitable automotive safety glass samples.
Materials Science Research: Spectral characterization of materials across ultraviolet, visible and near-infrared regions.
Scientific Research: Optical measurement and spectral analysis in research laboratories.
Higher Education: Spectroscopy experiments and material analysis in academic laboratories.